The most realistic information about the transparent sample such as a live cell can be obtained only using bright-field light microscopy. At high-intensity pulsing LED illumination, we captured a primary 12-bit-per-channel (bpc) response from an observed sample using a bright-field microscope equipped with a high-resolution (4872x3248) image sensor. In order to suppress data distortions originating from the light interactions with elements in the optical path, poor sensor reproduction (geometrical defects of the camera sensor and some peculiarities of sensor sensitivity), we propose a spectroscopic approach for the correction of this uncompressed 12-bpc data by simultaneous calibration of all parts of the experimental arrangement. Moreover, the final intensities of the corrected images are proportional to the photon fluxes detected by a camera sensor. It can be visualized in 8-bpc intensity depth after the Least Information Loss compression [Lect. Notes Bioinform. 9656, 527 (2016)].
翻译:在高强度脉冲LED光照下,我们用一个配备高分辨率(4872x3248)图像传感器的光场显微镜,从观测的样品中采集到初级12位/每声道(bpc)反应。为了抑制光场显微镜与光路元素、低传感器复制(相机传感器的几何缺陷和传感器敏感度的某些特殊性)的光谱相互作用所产生的数据扭曲,我们建议一种光谱方法,通过同时校准实验安排的所有部分来纠正这一未压缩的12位/每声道数据。此外,经过校正的图像的最后强度与摄像传感器检测到的光通量成比例。在最小信息损失压缩[Lect. Notesform.9656, 527-2016]之后,可以在8位/bc强度深度的深度内进行视觉化[Lect. notephicform.9656, 527-107]。